SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X MM56; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
71332756 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Post Operative Wound Infection (2446)
|
| Date of Event |
10/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed due to an infection, 2 months post total hip arthroplasty.The joint was washed out, and the poly was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a revision surgery was performed due to an infection 2 months post total hip arthroplasty.The joint was washed out and the poly and the femoral head were explanted.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode.The device was sterilized according to sterilization release documentation from quality control.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.A clinical evaluation noted that no relevant supporting documentation was provided, therefore a thorough medical investigation could not be performed.Based on the limited information available, the root cause of the infection 2 months post-op cannot be determined at this time.Without the return of the actual product involved and no medical records available, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|