MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number FLOW-I C20
Device Problem Output Problem (3005)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2020
Event Type  malfunction  
Event Description
It was reported that while the anesthesia workstation was in use, alarms for high airway pressure were generated.There was no patient injury.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
The anesthesia system was investigated by our field service engineer on site.A system checkout was performed which successfully passed and no abnormalities were found during ventilation.No part needed to be replaced.A complete set of device logs was saved and sent for evaluation.The evaluation of the provided logs confirms the reported alarm and it is likely that there was a leakage that caused self triggering which in turn led to high inspiratory flows and thus causing the generation of an airway pressure alarm followed by a pressure relief by opening the fresh gas safety valve.
 
Event Description
Manufacturers ref: # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key10902469
MDR Text Key218244176
Report Number8010042-2020-01066
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K191027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLOW-I C20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received11/25/2020
Supplement Dates Manufacturer Received01/25/2024
Supplement Dates FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-