MAUDE Adverse Event Report: EPOCAL INC. EPOC READER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EPOCAL INC. EPOC READER Back to Search Results
Catalog Number 10736398
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens has completed the investigation: the cause of the hematocrit (hct) discrepancy could not be determined from the information provided by the customer.A review of the in-house performance (including retain testing) for the card lot used, 10-20216-20, did not identify any product deficiencies.The failure rate of lot 10-20216-20 is not showing an increased trend in the field, therefore, there is no further evidence that the system or reagent cards are not performing as intended.The csv files provided by the customer offered additional insights and brought value to the investigation.From the information provided by the customer, there are a few factors that may have contributed to a discrepant hct reading.Firstly, different sample draws (collection site, collection method, time delays), sample handling and testing delays are known to affect test results if proper precautions are not considered.On 12 nov 2020, it was indicated that the epoc sample was tested at 5:41 am, resulting in an iqc failure (sample delivery failure).The same sample was then tested again at 5:47 am on epoc (6 min delay), providing the hct result of 27 %pcv.It should be noted that delays in testing a drawn blood sample in the absence of anticoagulant increase the risk associated with clotting and mixing (sample handling) related issues - which in turn can lead to discrepant readings.Secondly, the methods of hct measurement are not the same with epoc and the comparative instrument.Sysmex uses a complete blood cell count (cbc) method, which has known limitations associated with the calculation of falsely elevated values when a high reticulocyte or white blood cell (wbc) count is present.This is because the higher cell volumes of wbcs/ reticulocytes can interfere with the rbc count and thus falsely increase the hct reading.The customer has confirmed the patient's reticulocyte count was slightly elevated and wbc count was high, therefore, the impact/contribution of this measurement method cannot be ruled out as a potential elevated result on the sysmex.Thirdly, it was indicated by the customer that the sysmex xn 3000 utilized a sample with k2 edta anticoagulant.It is well documented in literature that the mean cell volumes of k3 edta anticoagulated blood are approximately 4% less than k2 edta anticoagulated blood.While the choice of anticoagulant affects the microhematocrit method to which all hematocrit methods are calibrated, results from routine samples on hematology analyzers are independent of the anticoagulant used.Since most clinical hematology analyzers are calibrated by the microhematocrit method using k3 edta anticoagulant, the epoc system is calibrated to a k3 edta microhematocrit method.If the sysmex instrument is calibrated to k2 edta anticoagulated blood, it will read ~4% higher than epoc.In these situations epoc advises a correction factor of 1.045 (e.G.K2 edta sysmex / 1.045).
 
Event or Problem Description
The customer reported a discrepant hematocrit result on a neonate on the epoc reader when compared to the result on a non-siemens hematology instrument.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPOC READER
Common Device Name
EPOC
Manufacturer (Section D)
EPOCAL INC.
2060 walkley road
ottawa, ontario K1G 3 P5
CA  K1G 3P5
MDR Report Key10905601
Report Number3002637618-2020-00067
Device Sequence Number3045934
Product Code CGL
UDI-Device Identifier00809708121860
UDI-Public00809708121860
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K113726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number10736398
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/12/2020
Initial Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-