| Model Number |
45-480420-L |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for evaluation.The investigation is currently underway.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a peripheral coil embolization procedure, the embolization coil would not advance and then detached during the removal attempt.The coil was removed in its entirety together with the catheter.There was no reported patient injury or additional intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned device confirmed the implant coil was detached from the pusher and stuck inside the microcatheter, as reported.The exact conditions or circumstances that led to the detachment could not be determined; however, the findings are consistent with damage that occurs from the application of excessive forces to the monofilament.
|
| |
|