| Model Number |
FX-410615 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and exchange from textured to smooth implants.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|