|
Inspection results of the returned device identified three kinks in the distal catheter section at 5.7cm, 7.5cm, and 14cm from the distal tip.Inspection through the inner diameter of the catheter identified resistance when verifying the internal diameter at 14cm from the distal tip with minor delamination visible which is indicative of a kinked shaft.No inner damage was observed at kink locations of 5.7cm and 7.5cm from the distal tip.The returned device marker band displayed severe ovalization, and it is uncertain if it was caused during the case, or subsequent device handling and transportation.If zoom 88t was kinked and the intermediate catheter was continuously pulled against the resulting resistance, the intermediate catheter may stretch and detach.The returned device marker band displayed severe ovalization and is likely a result of the handling and transportation of the returned device, because during the case the zoom71 and subsequent competitor devices were able to exit the zoom 88t tip without issues.The root cause of resistance was unable to be determined as the provided case images did not capture the pertinent vessel anatomy at moment of detachment.Per instructions for use, the devices should not be withdrawn against resistance without careful assessment of the cause under fluoroscopy.If the cause cannot be determined, withdraw all devices as a single unit.The manufacturing records of this lot were reviewed and did not reveal any issues pertaining to design, manufacturing, or quality.Appropriate testing and inspection was completed to ensure the device met minimum tensile specification and is kink resistant.The distal section undergoes 100% visual inspection and is free of visual defects or protrusions.The lot number was not provided by the user facility.All devices are required to meet lot release testing, which includes kink resistance of the distal section.Catheters are 100% inspected to be free of visual defects or kinks during in-process inspection and during quality inspection after manufacturing.This report is associated with mfg report number: 3014590708-2020-00006.
|