|
Inspection results of the returned device confirmed that the distal section of the catheter was separated.The returned device detached at 0.4 cm from the distal tip, and the remaining shaft measured at 137.8cm.The detached portion includes a slightly folded marker band, stretched jacket material, and coil.A kink was present at 15.0cm from the tip of the remaining shaft.It is uncertain if this kink was present during the procedure or was a result of handling damage after removal from the patient.Based on provided case information, resistance was encountered, and the zoom 71 continued to be retracted against resistance.If the device is constrained and forcefully retracted against resistance, the device may stretch and detach.The root cause of the resistance and detachment is unable to be determined as the provided case images did not capture the pertinent vessel anatomy at moment of detachment.Per instructions for use, the devices should not be withdrawn against resistance without careful assessment of the cause under fluoroscopy.If the cause cannot be determined, withdraw all devices as a single unit.The manufacturing records of this lot were reviewed and did not reveal any issues pertaining to design, manufacturing, or quality.Appropriate testing and inspection was completed to ensure the device met minimum tensile specification and is kink resistant.The distal section undergoes 100% visual inspection and is free of visual defects or protrusions.This report is associated with mfr report number: 3014590708-2020-00007.
|