(b)(4).Batch #: u5a025.Investigation summary: the analysis results found that the ats45nk device was received with no apparent damaged and with a tr45w cartridge loaded on the device.The returned reload was partially fired 1/3.The returned device was tested for functionality with a test reload and it fired, and all the staples formed as intended.The staple line was complete and the staples meet the staple form release criteria.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.
|