Model Number ATS45NK
Medical Device Problem Codes Failure to Fire (2610); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: u5a025.Investigation summary: the analysis results found that the ats45nk device was received with no apparent damaged and with a tr45w cartridge loaded on the device.The returned reload was partially fired 1/3.The returned device was tested for functionality with a test reload and it fired, and all the staples formed as intended.The staple line was complete and the staples meet the staple form release criteria.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.
 
Event or Problem Description
It was reported that the ats45nk device didn¿t work as expected.On (b)(6) prof was using this in a hepatectomy case and the second click didn¿t appear to work properly.The surgeon struggled to remove the device from the patient and it hadn¿t fired.The patient was unharmed and the case continued with the surgeon using another device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS ARTIC LNR CUTR NO KNIFE
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key10910971
Report Number3005075853-2020-06262
Device Sequence Number3317217
Product Code GDW
UDI-Device Identifier10705036001089
UDI-Public10705036001089
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2024
Device Model NumberATS45NK
Device Catalogue NumberATS45NK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/29/2020
Initial Report FDA Received Date11/26/2020
Patient Sequence Number1
-
-