MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX22508UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 11/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure, an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a severely tortuous, moderately calcified lesion located in the proximal circumflex (cx) artery.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during delivery to/at a lesion.The stent was successfully removed from the patient using a snare.It was stated that tangled wired and use of guide extension catheter believed by physician to have caused dislodgment and there is no negative product accusations.The patient was reported to be alive with no further injury.
 
Additional Manufacturer Narrative
The device was prepped appropriately with no issues noted.No resistance was noted during withdrawal of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key10911222
Report Number9612164-2020-04639
Device Sequence Number3170422
Product Code NIQ
UDI-Device Identifier00643169556690
UDI-Public00643169556690
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2022
Device Model NumberRONYX22508UX
Device Catalogue NumberRONYX22508UX
Device Lot Number0010211916
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/23/2020
Supplement Date Received by Manufacturer12/07/2020
01/06/2021
Initial Report FDA Received Date11/26/2020
Supplement Report FDA Received Date12/15/2020
01/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-