SMITH & NEPHEW, INC. TWNFX TI 3.5 STR ANCHR W/2 ULTRB STR N; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
7210707 |
| Medical Device Problem Codes |
Break (1069); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Foreign zip code: (b)(6).Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a foot procedure, the needles came off from the twinfix anchor while tying the knots and or passing.Also some suture breakage was experienced.It is unknown how the procedure was completed.No significant delay was reported.No further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Internal complaint reference: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|