SMITH & NEPHEW, INC. JRNY II CR ISRT TRL RT SZ 5-6 9M; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74033651 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a total knee arthroplasty, the insert trial cracked.The insert trial broke after the femur and tibia had already been cemented.Small piece was stuck in the back of the knee.Surgeon was able to retrieve the small broken piece from the back of the knee.The procedure was finished using a smith and nephew backup device.No delay or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken and has signs of wear and tear from use.The device was manufactured in 2015.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of the risk management file revealed this failure mode was previously identified.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|