• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OREGON MANUFACTURING PLANT NATURALYTE 2301 2K 3.0CA 1MG GAL; DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OREGON MANUFACTURING PLANT NATURALYTE 2301 2K 3.0CA 1MG GAL; DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER) Back to Search Results
Model Number NATURALYTE 2301 2K 3.0CA 1MG GAL
Medical Device Problem Code Chemical Problem (2893)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A fresenius sales manager (fsm)reported to fresenius customer service that a hemodialysis (hd) clinic reported out of range conductivity levels with naturalyte during testing/set-up prior to initiating treatment.There was no patient involvement reported.Upon follow up, the fsm reported during setup the conductivity was at 13.1 ms/cm and the theoretical conductivity value (tcd) was set to 13.7 ms/cm.He confirmed no patients were treated with this batch.There was no recent service to the machine.There were a total of 47 cases of this reported lot.The fsm was informed a pick up will be scheduled for a sample return.They are currently using naturalyte bicarbonate, product and lot number are unknown.
 
Additional Manufacturer Narrative
Plant investigation: a product history review was performed.A review of the complaint management system identified 5 additional complaints related to conductivity issues with the reported lot.A review of the product inventory records for lot no.Indicated that (b)(6) of finished product were manufactured for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac057 met release specifications.As of 12/17/2020 no samples were returned.Samples are not needed to confirm the allegation.Through other complaint allegations of the same issue for lots 20lxac056, 20lxac058, and 20lxac076, there were companion samples available for testing which were tested at both the oregon and irving laboratories and results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac057 did not meet release specifications and the allegation conductivity low was confirmed.In conclusion, 20lxac057 release testing was found to be compromised due to the deficient test method.A non-conformance was opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURALYTE 2301 2K 3.0CA 1MG GAL
Common Device Name
DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)
Manufacturer (Section D)
OREGON MANUFACTURING PLANT
director, quality systems
750 n lallendorf road
oregon OH 43616
MDR Report Key10916188
Report Number3005162618-2020-00049
Device Sequence Number4258481
Product Code KPO
UDI-Device Identifier00840861101702
UDI-Public00840861101702
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K192017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model NumberNATURALYTE 2301 2K 3.0CA 1MG GAL
Device Catalogue Number08-2301-3
Device Lot Number20LXAC057
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Received by Manufacturer 11/06/2020
Supplement Date Received by Manufacturer12/22/2020
Initial Report FDA Received Date11/29/2020
Supplement Report FDA Received Date12/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-