• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 LCS HP FEM NOTCH IMPACTOR; KNEE INSTRUMENT : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 LCS HP FEM NOTCH IMPACTOR; KNEE INSTRUMENT : IMPACTORS Back to Search Results
Model Number 9505-01-219
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
As sterile processed loan trays for lcs tkr were being opened in theatre, the scrub nurse noticed a white substance on the tibial and femoral impactor.The substance appeared to be cement from previous use.On inspection of other trays by rep (myself), it was noticed that some instrumentation was not disassembled from previous cases.The surgeon was made aware of the contamination and the surgery was cancelled.The surgeon spoke with the patient and rebooked the case for two weeks time.
 
Event or Problem Description
1.Kindly clarify why some instrumentation was not disassembled, was there any product issue of these products? ¿ where appropriate instruments that require it are disassembled during the inspection process.In these specific cases the instruments are then reassembled and placed back into the tray as this is how they fit into the trays.They are not sent to customers disassembled.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS HP FEM NOTCH IMPACTOR
Common Device Name
KNEE INSTRUMENT : IMPACTORS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10919343
Report Number1818910-2020-25619
Device Sequence Number16812152
Product Code HWA
UDI-Device Identifier10603295225263
UDI-Public10603295225263
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9505-01-219
Device Catalogue Number950501219
Device Lot NumberPO0508
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/12/2020
Supplement Date Received by Manufacturer12/01/2020
02/03/2021
Initial Report FDA Received Date11/30/2020
Supplement Report FDA Received Date12/11/2020
02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-