Model Number 71365600
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that sizing issues with the anthology case were noticed.It was not possible to attribute the issue to any particular instrument and there is no singular surgery to cite.No further information reported.
 
Additional Manufacturer Narrative
Investigation results: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device found the device to be dull.The device show signs of extensive use.The dimensional evaluation could not verify the failure mode.The returned devices were verified to be within drawing print specifications and tolerances.The broaches were verified to be with the appropriate overlays for the tooth profile.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY STARTER BROACH
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10920264
Report Number1020279-2020-06863
Device Sequence Number11382273
Product Code JDG
UDI-Device Identifier03596010559807
UDI-Public03596010559807
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71365600
Device Catalogue Number71365600
Device Lot Number13MM11479
Initial Date Received by Manufacturer 11/03/2020
Supplement Date Received by Manufacturer03/16/2021
Initial Report FDA Received Date11/30/2020
Supplement Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-