| Model Number |
4-840120DIUU-US |
| Medical Device Problem Codes |
No Flow (2991); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that while in use on a patient, the 840 ventilator powered off automatically.The patient was placed on an alternate ventilator without harm.
|
| |
|
Event or Problem Description
|
|
Correction: it was initially reported that while in use on a patient, the 840 ventilator powered off automatically.The patient was placed on an alternate ventilator without harm.Additional information received from the rental company¿s biomed manager, that the ventilator did not stop delivering breaths to the patient based on their evaluation of the ventilator and the ventilator logs,.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to section h6 device codes and evaluation code result.Additional code added to section h6 evaluation code method.Additional information received as follows: the third party service technician evaluated the ventilator and could not duplicate the reported event.The ventilator passed all tests, no parts required replacement, and the preventative maintenances were up-to-date.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|