• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 4-840120DIUU-US
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient, the 840 ventilator stopped ventilating and generated a breath delivery unit (bdu) ba ckground check code which indicated that the inspiratory pressure transducer was stuck.The patient was removed from the ventilator and placed on an alternate ventilator without harm as the result of the event.
 
Additional Manufacturer Narrative
Correction to section h6 evaluation code method and result.Component code added.H3 device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that while in use on a patient, the 840 ve ntilator generated a breath delivery unit (bdu) background check code which indicated that the inspiratory pressure transducer was stuck.The device wass available for evaluation.The service personnel (sp) inspected the ventilator and confirmed the reported issue.To solve the issue, sp replaced ai (analog interface) pcba (printed circuit board assembly) and verified latest software and fpga¿s (field programmable gate array).The ventilator passed all tests and calibrations per manufacturer specifications at the time of service and was returned to the customer.One ai pcb was returned for failure investigation.The component was visually inspected and functionally tested with no malfunction or product deficiency identified.The event was isolated in the field to a potential fault with ai pcba; however, the returned component was analyzed and no fault was found.With the information available a likely cause could not be determined.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality s pecifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
MDR Report Key10920439
Report Number8020893-2020-00188
Device Sequence Number11452675
Product Code CBK
UDI-Device Identifier10884522019354
UDI-Public10884522019354
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K970460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4-840120DIUU-US
Device Catalogue Number4-840120DIUU-US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Initial Date Received by Manufacturer 11/05/2020
Supplement Date Received by Manufacturer09/16/2021
Initial Report FDA Received Date11/30/2020
Supplement Report FDA Received Date09/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
Patient Weight117
-
-