Plant investigation: the customer reported that samples were available for return, however as of 11/12/2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications.In addition, there were companion samples available for testing which were tested at both the oregon and irving laboratories, and results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac076 did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/19/2020 did not identify any additional reported events for lot no.20lxac076.A review of the product inventory records for lot no.20lxac076 indicated that 4507 cases of finished product were manufactured on 9/20/20 for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac076 met release specifications.In conclusion, 20lxac076 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) area technical operations manager (atom) reported to fresenius customer service that a lot of naturalyte had low conductivity.Upon follow up, the atom reported that before use, the conductivity was at 13.1ms/cm.The theoretical conductivity value (tcd) setting on the machine was not provided and the difference between actual conductivity is unknown.No patients were treated with the lot.The atom stated they used one jug from the lot, however they had eight and a half cases that would not be used.The atom reported that they use bibags for bicarbonate.A return goods authorization (rga) was issued to the clinic for product return.
|