Plant investigation: the customer reported that samples were available for return, however as of 11/12/2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications.In addition, there were companion samples available for testing which were tested at both the (b)(6) laboratories, and results were found to be conflicting.Due to the conflicting results found between the laboratories, the (b)(6) test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac056 did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/13/2020 identified 14 additional reported events for lot no.20lxac056 related to conductivity issues.The threshold for non-conformance was reached for lot 20lxac056, however a non-conformance was already initiated for this issue.A review of the product inventory records for lot no.20lxac056 indicated that (b)(4) cases of finished product were manufactured on 9/16/20 for distribution with no additional noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac056 met release specifications.In conclusion, 20lxac056 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) clinic biomedical technician (biomed) reported to fresenius customer service that a lot of naturalyte had low conductivity.Upon follow up, the biomed reported that the conductivity was at 13.1 ms/cm.The theoretical conductivity value (tcd) setting on the machine was not provided and the difference between actual conductivity is unknown.No patients were treated with the lot.Per the biomed, six jugs were affected, and seven cases were set aside.The biomed reported that there was no service to the machines and that they use centrisol for bicarbonate.The biomed reported that a sample was available for return to the manufacturer for physical evaluation.
|