Model Number 11327700
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported event.A supplement report will be filed upon completion of the investigation.Internal id# (b)(4).
 
Event or Problem Description
It was reported to siemens that during an emergency procedure on the artis icono floor unit an x-ray image was pixelated.The image was unusable for diagnosis, and the procedure had to be completed on an alternate system.The reported event occurred in (b)(6).
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The returned x-ray tube was examined at the factory and the small focus filament was found to be defective.Under normal conditions, each filament (emitter/heating element) has the same distance to the boundary plates surrounding it.However, thermal stress can have a negative effect on the position of the filament and the distance can decrease on one side.In the present case, the emitter was so close to the outer focal boundary that it touched the focal head, resulting in the unavailability of small focus x-rays.An x-ray tube is a wear part and filaments are the typical wear parts that limit the life of the tube.In such a case, the system will automatically switch to the next larger focus after pressing the foot switch three times as described in the user manual, resulting in a moderate degradation of image quality (e.G.Resolution / sharpness).Generally, acceptable residual functionality remains, yet in the given case the operator decided to finish the operation on another system.The affected x-ray tube has been exchanged on site by the local service organization and the error has not been reported again.The spare parts consumption was checked and a possible general fault that would require correction of the installed base could not be identified by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ICONO FLOOR
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, 91301
GM  91301
MDR Report Key10924598
Report Number3004977335-2020-57139
Device Sequence Number16383684
Product Code OWB
UDI-Device Identifier04056869149325
UDI-Public04056869149325
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K193326
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11327700
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/24/2020
Supplement Date Received by Manufacturer02/10/2021
Initial Report FDA Received Date11/30/2020
Supplement Report FDA Received Date02/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-