AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER II; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA2-06-06
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Codes Embolism (1829); Foreign Body Embolism (4439)
Date of Event 11/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
On (b)(6) 2020, a 6/6 amplatzer duct occluder ii was selected based on measurements of the ductal anatomy for implant in the patient.During the procedure the device was positioned, released from the delivery cable and then embolized into the patient's pulmonary trunk.The device was retrieved using a merit medical snare.The patient did not experience any serious injuries and a 6mm amplatzer duct occluder(ado) was used in replacement since the ductus was too lax.The ado device was successfully implanted in the patient.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay in the procedure.The patient is currently stable.
 
Additional Manufacturer Narrative
An event of the device embolizing into the patient's pulmonary trunk was reported.The investigation confirmed the device met dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER II
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10924745
Report Number2135147-2020-00512
Device Sequence Number959915
Product Code MAE
UDI-Device Identifier00811806011264
UDI-Public00811806011264
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number9-PDA2-06-06
Device Catalogue Number9-PDA2-06-06
Device Lot Number7117402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/04/2020
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient Weight59
-
-