• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVOS; OXIMETER, TISSUE SATURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number IS
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Pressure Sore/Ulcer (2326)
Date of Event 11/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, there was a wound below the cerebral sensor on a premature patient.The patient was injured.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the sensor had been left on the premature patient for over four days.It was reported that there was a wound below the cerebral sensor on a premature patient.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: use care when placing or removing sensor.Do not place on broken or undeveloped skin.To avoid pressure sores do not apply pressure (e.G.Headbands, wraps, tape) to sensor and periodically check skin integrity according to your institution¿s patient care protocol or at least every 24 hours.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
MDR Report Key10924746
Report Number2936999-2020-00901
Device Sequence Number11424193
Product Code MUD
UDI-Device Identifier10884521211117
UDI-Public10884521211117
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K091224
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIS
Device Catalogue NumberIS
Device Lot NumberIS
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/09/2020
Supplement Date Received by Manufacturer10/14/2021
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date10/19/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age7 DA
Patient Weight1
-
-