|
Plant investigation: the customer reported that samples were available for return, however as of: 11/12/2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications through other complaint allegations of the same issue for lots:20lxac056, 20lxac058, and 20lxac076.There were companion samples available for testing which were tested at both the oregon and irving laboratories.And results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot: 20lxac090.Did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/13/2020 identified 3 additional reported events for lot no: 20lxac090 related to conductivity issues.A review of the product inventory records for lot no: 20lxac090, indicated that (b)(4) cases of finished product were manufactured on 9/22/20 for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac090 met release specifications.In conclusion, 20lxac090 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) clinic biomedical technician (biomed) reported to fresenius customer service that a lot of naturalyte had low conductivity.Upon follow up, the biomed reported that during a patient¿s treatment, the conductivity was at 13.1 ms/cm and fluctuating.The theoretical conductivity value (tcd) setting on the machine was not provided and the difference between actual conductivity is unknown.The patient completed treatment on the machine with the same jug of naturalyte.The patient did not experience any adverse events.Per the biomed, they used other jugs from the same lot without any issues.The biomed reported that there was no service to the machines and that they use bicarbonate 4000rx.The sample was discarded and is not available for return to the manufacturer for physical evaluation.
|