Plant investigation: as of (b)(6) 2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications.In addition, there were companion samples available for testing which were tested at both the oregon and irving laboratories, and results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac056 did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/13/2020 identified 13 additional reported events for lot no.20lxac056 related to conductivity issues.The threshold for non-conformance was reached for lot 20lxac056, however a non-conformance was already initiated for this issue.A review of the product inventory records for lot no.20lxac056 indicated that (b)(4) cases of finished product were manufactured on 9/16/20 for distribution with no additional noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac056 met release specifications.In conclusion, 20lxac076 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) clinic biomedical technician (biomed) reported that some naturalyte had low conductivity and four cases were pulled from the shelf.Upon follow up, the customer said that the conductivity was at 13.2-13.3 ms/cm and the machine alarmed low conductivity.The biomed stated the product was within the acceptable range, so it was used for patient treatment without issues.There was no patient adverse event.A return goods authorization (rga) was issued to the clinic for product return.
|