• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS TORIC II 1-PIECE ACRYLIC; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS TORIC II 1-PIECE ACRYLIC; TORIC IOLS Back to Search Results
Model Number ZCU525
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted, give date: not applicable, as lens remains implanted.Device evaluation: product evaluation was not performed because per the initial report the lens is still implanted in the patient's eye.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) had rotated approximately 11 degrees post implant.The outcome does not significantly interfere with the patient¿s daily activities.It was stated that the initial operative iol axis estimate was for the right eye (od) 087 degrees.However, 3-month postoperative iol axis estimate od was 076 degrees.The uncorrected visual acuity od at the 3-month postoperative visit: 20/25.Manifest refraction od: +0.25 sph.Best-corrected visual acuity od: 20/20.There was no reported visual symptoms at this visit.No surgical intervention is planned at this time.No other information was provided.
 
Additional Manufacturer Narrative
Additional information: the complaint was based upon a subjective estimate made by the investigator from a slit-lamp measurement.For this study, we have also been doing independent analysis of the intraocular lens (iol) photos taken at each visit.These are objective measurements of the actual iol axis.Based on this objective measurement, the actual amount of iol rotation for the subject below is less than 01 degrees.There is still no surgical intervention planned, and the subject has exited the study.Subject visit schedule page study eye delta angle from base: 80729512008 po 3 month analyst 1 review od 0.714; 80729512008 po 3 month analyst 2 review od 0.629.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS TORIC II 1-PIECE ACRYLIC
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10926165
Report Number9614546-2020-00504
Device Sequence Number11815804
Product Code HQL
UDI-Device Identifier05050474664210
UDI-Public(01)05050474664210(17)241009
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2024
Device Model NumberZCU525
Device Catalogue NumberZCU525U165
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/02/2020
Supplement Date Received by Manufacturer01/22/2021
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date02/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-