An angiodynamics territory manager reported an issue with an evlt kit.During preparation, it was noted that the fiber was broken, upon opening.The device was set aside and the procedure was completed with another same device and there was no report of patient harm or adverse event, by the end user.It was indicated that the reported device is not available to be returned to the manufacturer for evaluation.It was indicated the reported device is available for return to the manufacturer for a device evaluation.
|
|
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint of fiber broken/fractured cannot be confirmed as no product was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A review of the incoming lot history records was performed for the reported packaging lots for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.During the packaging step, the fibers are inspected for damage.It is unlikely that the fiber was fractured prior to packaging.The most likely root cause to this event is due to handling after the device left the angiodynamics facility.Labeling review: the directions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference(b)(4).
|