Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that two (2) screws fractured and there was a sixty (60) minute delay while the fractured shafts of the screws were removed from the patient.The patient did not have a catheter at the beginning of the surgery, and a catheter was put in after the screw fractures as the surgeon realized the surgery would take longer than anticipated.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event or Problem Description
It was reported that two (2) screws fractured and there was a sixty (60) minute delay while the fractured shafts of the screws were removed from the patient.The patient did not have a catheter at the beginning of the surgery, and a catheter was put in after the screw fractures as the surgeon realized the surgery would take longer than anticipated.The screws were removed by cutting away bone and tissue around the broken screw in order to make enough room for wire drivers to grab onto the end and twist it out.No additional patient consequences have been reported.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A visual inspections of the returned screws show heavy signs of use.The screws have both fractured where the screw threads meet the screw head.The complaint is confirmed.A determination cannot be made as to what caused the screws to fracture.Lot identification is necessary for review of device history records, lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMAONE SYSTEM 1.7MM X 7MM TI SELF DRILLING IMF SCREW
Common Device Name
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key10927151
Report Number0001032347-2020-00593
Device Sequence Number1429630
Product Code DZL
UDI-Device Identifier00841036194390
UDI-Public00841036194390
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K983728
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberSP-3236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/16/2020
Supplement Date Received by Manufacturer11/24/2020
04/06/2021
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date12/17/2020
04/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-