| Model Number |
10067 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Coopersurgical , inc.Is currently investigating the reported condition.
|
| |
|
Event or Problem Description
|
|
Description of event: "md attempting vacuum.Both items broke at the junction where the mushroom cap meets the long plastic arm.Physician questions if the pop off mechanism was working as she felt it should have popped off instead of break." mityone mushroom cup 10067 (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Review dhr inspect returned samplesinvestigation initiated manufacturer's investigation.Distribution history the complaint product was manufactured at csi on 07-14-2020 under work order (b)(4).Manufacturing record review dhr-10067 - 290007 was reviewed and no non-conformities, related to the complaint condition, were noted.Incoming inspection review incoming inspection record review not applicable to this product.Service history record service history not applicable for this product.Historical complaint review a review of the 2-year complaint history showed similar reported complaint conditions.Product receipt the complaint product was returned to csi and confirmed 1/14/2021.Visual evaluation a visual evaluation of the complaint product was completed and confirmed 1/14/2021.Functional evaluation an evaluation of the complaint product could not be completed as the complaint product was returned in a nonfunctional manner due to reported event, see attached email with photos.Root cause definitive root cause as to the resulting reported event is indeterminable.Pull test supplier data indicated acceptable results.Corrective actions not applicable as the affected devices were individually tested before packaging and were found to be acceptable.No further action is required at this time.Was the complaint confirmed? yes.
|
| |
|
Event or Problem Description
|
|
Report stated: "md attempting vacuum.Both items broke at the junction where the mushroom cap meets the long plastic arm".10067 mityone mushroom cup (b)(4).
|
| |
|
Search Alerts/Recalls
|