ANGIODYNAMICS BIOFLO DIALYSIS CATHETER; CATHETER, HEMODIALYSIS, IMPLANTED Back to Search Results
Catalog Number 10303803
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 11/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the disposable device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
A territory manager reported the end user had a problem with a bioflo duramax dialysis catheter.During placement of the bioflo duramax, the dialysis catheter detached from the tunneler and the sleeve broke off the tunneler, becoming stuck inside the patient.The catheter sleeve was removed with no adverse patient effect due to this event.It was indicated that the reported device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of tunneler/sleeve/catheter connection issue (loose/detached) cannot be confirmed, no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.The tunneler sleeve accessory is supplied to angiodynamics by the supplier martech.Scar004238 was sent to martech for dhr review of affected supplier lot.As per scar004238 results from martech, the dhr review showed no issues with the production paperwork.The tri-ball tunneler accessory is supplied to angiodynamics by the supplier disposable instrument.Scar004227 sent to the supplier, disposable instrument for dhr review of supplier lot.As per scar004227 results from disposable instruments.The dhr review showed no issues with the production paperwork.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the following is provided as a reference from dfu: warning: do not over-expand subcutaneous tissue during tunneling.Over-expansion may delay/prevent cuff in-growth.6.Lead catheter into the tunnel gently.Do not pull or tug the catheter tubing.If resistance is encountered, further blunt dissection may facilitate insertion.Remove the catheter from the trocar with a slight twisting motion to avoid damage to the catheter.Precaution: do not pull tunneler out at an angle.Keep tunneler straight to prevent damage to catheter tip.Note: a tunnel with a wide gentle arc lessens the risk of kinking.The tunnel should be short enough to keep the y-hub of the catheter from entering the exit site, yet long enough to keep the cuff 2 cm (minimum) from the skin opening.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO DIALYSIS CATHETER
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
MDR Report Key10928007
Report Number1317056-2020-00193
Device Sequence Number716152
Product Code MSD
UDI-Device IdentifierH965103038031
UDI-PublicH965103038031
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K131260
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number10303803
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2020
Supplement Date Received by Manufacturer11/13/2020
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-