| Model Number |
20 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device would not complete the boot up process.There was no patient use reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Physio-control further evaluated the removed pcb assembly and determined that the cause of the reported issue was due to shorted capacitor, u61.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device would not complete the boot up process.There was no patient use reported with the event.
|
| |
|