Plant investigation: the customer reported samples were available for return for evaluation, however as of 11/12/2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications.Through other complaint allegations of the same issue for lots 20lxac056, 20lxac058, and 20lxac076, there were companion samples available for testing which were tested at both the oregon and irving laboratories.And results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac062 did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/19/2020 identified 1 additional reported event for lot no.20lxac062 related to conductivity issues.A review of the product inventory records for lot no.20lxac062 indicated that (b)(4) cases of finished product were manufactured on 9/16/20 for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac062 met release specifications.In conclusion, 20lxac062 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) clinic biomedical technician(biomed) reported to fresenius customer service that a lot naturalyte has low conductivity and specific gravity.Upon follow up, the biomed reported that before use, the conductivity was at 13.0.No patients were treated with the lot.Per the biomed, 2 cases are affected and now they are using naturalyte lot 20kxac099 without any issues.The biomed reported that there was no service to the machines and that they use bicarbonate naturalyte 45x, lot number 20lxbc010.
|