• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 4 10MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 4 10MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-50-410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Medical records received and the patient had a second revision of her right knee for pain.All implants were revised - except the patella.Competitor products were implanted.Doi: (b)(6) 2016 (tibial component); doi: (b)(6) 2017 (tibial insert, femoral component, and cement); dor: (b)(6) 2020; right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS RP INSRT SZ 4 10MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10929052
Report Number1818910-2020-25845
Device Sequence Number4169187
Product Code NJL
UDI-Device Identifier10603295052326
UDI-Public10603295052326
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2018
Device Model Number1516-50-410
Device Catalogue Number151650410
Device Lot Number4768205
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date12/01/2020
Supplement Report FDA Received Date01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE MEDIAL ANAT PAT 35MM; ATTUNE PS FEM RT SZ 4 CEM; ATTUNE PS RP INSRT SZ 4 10MM; ATTUNE RP TIB BASE SZ 4 CEM; SMARTSET GMV 40G US EO; SMARTSET GMV 40G US EO; ATTUNE MEDIAL ANAT PAT 35MM; ATTUNE PS FEM RT SZ 4 CEM; ATTUNE PS RP INSRT SZ 4 10MM; ATTUNE RP TIB BASE SZ 4 CEM; SMARTSET GMV 40G US EO; SMARTSET GMV 40G US EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-