Model Number 283512
Device Problems Mechanical Problem (1384); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be filed as appropriate.Investigation summary: the device was received and evaluated at the service center.The reported complaint that the device doesn't turn due to a jammed motor was confirmed.It was found that the motor did not run as it was corroded.Further, the resistance values of the keypad of the hand control set and the protective earth resistance of the motor cable were out of range and the buttons of the keypad were not functioning.The motor, motor cable and the hand control set were replaced and the device was repaired, tested and found to be fully functional.Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor.The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer.However, given the information provided, we cannot determine a definitive root cause for the defective keypad of the hand control set and the defective motor cable.A manufacturing record evaluation was performed for the finished device (serial number : (b)(4)), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi:( b)(4).
 
Event Description
It was reported by affiliate via phone that the micro tornado hp w handcontrol doesnt turn  jammed motor.The procedure was completed with a replacement.No delay.No patient consequences.The device is available to be returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10930041
MDR Text Key219330766
Report Number1221934-2020-03666
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2020
Initial Date Manufacturer Received 11/13/2020
Initial Date FDA Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2019
Is the Device Single Use? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-