Plant investigation: the customer reported samples were available for return for evaluation, however as of 11/12/2020 no samples were returned.Samples are not needed to confirm the allegation.A sample retain was tested through a special test request and found to be within specifications.Through other complaint allegations of the same issue for lots 20lxac056, 20lxac058, and 20lxac076, there were companion samples available for testing which were tested at both the (b)(6) laboratories.And results were found to be conflicting.Due to the conflicting results found between the laboratories, the (b)(6) test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac059 did not meet release specifications and the allegation conductivity low was confirmed.A product history review was performed.A review of the complaint management system on 11/23/2020 identified 4 additional reported event for lot no.20lxac059 related to conductivity issues.A review of the product inventory records for lot no.20lxac059 indicated that (b)(4) cases of finished product were manufactured on 9/17/20 for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac059 met release specifications.In conclusion, 20lxac059 release testing was found to be compromised due to the deficient test method.A non-conformance was already opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|
|
A hemodialysis (hd) clinic biomedical technician (biomed) reported technical support that the 2008t machine experienced low conductivity.Upon follow up, the biomed confirmed that the machine was returned to service without any repairs.The biomed reported that they experienced low conductivity while using naturalyte, lot 20lxac059.The biomed reported that before use, the conductivity was at 12.9.The patients that were treated with the lot did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.Per the biomed 9 cases of the product are affected and now they are using naturalyte lot 20lxac079 without any issues.The biomed reported that there was no service to the machines a return goods authorization (rga) was issued to the clinic for product return.
|