The event involved a 60¿ (152 cm) appx 1.9 ml, smallbore ext set w/microclave® clear, 0.2 micron low protein binding filter, spiros® w/red cap, clamp that the customer reported an unspecified chemotherapy medication leaking from the filter during patient use.There was patient involvement; however, no report of adverse event.This event captures the first of two incidents.
|
|
H10: one used list #ch3603, 60¿ (152 cm) appx 1.9 ml, smallbore ext sets w/microclave® clear, 0.2 micron low protein binding filter, spiros® w/red cap, clamp,( lot #unknown), one (1) new list #ch3603, 60¿ (152 cm) appx 1.9 ml, smallbore ext set w/microclave® clear, 0.2 micron low protein binding filter, spiros® w/red cap, clamp, (lot #5012557) and one (1) used bd 0.9% sodium chloride injection, usp 10 ml syringe were received and visually inspected.As received, the 0.2 filter vent material of the used set was slightly wetted out with prior infusate solution.No other damage or anomalies were seen.Each extension set was primed with water at gravity pressure and then leak tested according to product performance specifications.No leaks were observed.The reported complaint could not be replicated or confirmed.The lot history was reviewed and no nonconformities were found that may have contributed to the complaint.
|