| Model Number |
6248V-130 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant, while slitting the catheter, there was an issue slitting and the catheter completely broke in half.The sections were successfully removed from the patient and the procedure was completed without the use of another catheter.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary the delivery catheter was returned and analyzed.The analysis indicated that the catheter did not slit along the score lines.There was an issue with the catheter shaft.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.The mechanical operation of the catheter peeling/slitting/splitting was partially executed.Visual analysis of the delivery catheter indicated damage during use.The analyst noted the delivery catheter was returned without the dilator.The delivery catheter sheath was torn at 14 centimeters.Slitting did not start on the score line on the hub of the delivery catheter and did not continue on the score line at the distal end of the hub.Slitting terminated at 14.5 centimeters.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|