|
A biomedical technician (bmt) reported to fresenius customer service that a hemodialysis (hd) clinic experienced low conductivity with a batch of naturalyte during treatment.There was no reported adverse event, serious injury, nor required medical intervention related to the event.Upon follow up, the bmt reported during treatment the conductivity was at 13.1 ms/cm although the theoretical conductivity value (tcd) was set to 13.5.The naturalyte jug was switched out mid-treatment without any ill effects.There was no recent service performed on the machine.The bmt dumped out two 55 gallon drums of the same batch.No samples are available to return.They are currently using granuflo and bibags 4000rx 12 (unknown part and lot numbers).
|
|
Plant investigation: a product history review was performed.A review of the complaint management system identified no additional complaints related to conductivity issues with the reported lot.A review of the product inventory records for lot no.20lxac028 indicated that 144 drums of finished product were manufactured for distribution with no noted nonconformances.After reviewing the finished product release summary, it was determined that 20lxac028 met release specifications.As of 12/11/2020 no samples were returned.Samples are not needed to confirm the allegation.Through other complaint allegations of the same issue for lots 20lxac056, 20lxac058, and 20lxac076, there were companion samples available for testing which were tested at both the oregon and irving laboratories and results were found to be conflicting.Due to the conflicting results found between the laboratories, the oregon test methods were reviewed and a deficiency was found in the test method pertaining to the sodium and potassium analyte tests.Because of the deficiency found in the test method, it was determined that lot 20lxac028 did not meet release specifications and the allegation conductivity low was confirmed.In conclusion, 20lxac028 release testing was found to be compromised due to the deficient test method.A non-conformance was opened for allegations of conductivity issues and escalated to a corrective action preventative action (capa).Based on the investigation performed, cause was traced to manufacturing.
|