Model Number ZCT600
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Date of Event 10/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluation: product evaluation was not performed because per the initial report the lens was discarded by the customer.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.Note: the device was manufactured at the kulim site which has been closed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis toric intraocular lens (iol) had rotated 24 degrees after one week postop with manifest refraction (mrx) >1.6 spherical equivalent.Patient was 20/20 uncorrected at post-op day 1.At postop week 1, her vision was 20/100.The surgeon planned on rotation only, but the patient was -1.5 so the surgeon decided on a lens exchange with a lower diopter.However, the ocular response analyzer (ora) stated to use the same diopter 16.0 again.The patient also experienced blurry vision.The lens was explanted and replaced with the same model and diopter.The explanted lens was discarded.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10933213
Report Number2020664-2020-00056
Device Sequence Number2709325
Product Code HQL
UDI-Device Identifier05050474604087
UDI-Public(01)05050474604087(17)221026
Combination Product (Y/N)Y
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/26/2022
Device Model NumberZCT600
Device Catalogue NumberZCT600U160
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2020
Initial Report FDA Received Date12/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-