• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC SIMPLY LEGACY 2 IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC SIMPLY LEGACY 2 IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 825713U
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation results are not available.When the analysis is complete, a supplemental report will be submitted.This complaint is being submitted late due to the furloughs that resulted from the global pandemic and is captured within (b)(4).
 
Event or Problem Description
Per complaint (b)(4), during clinical procedure, patient experienced lack of primary stability.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLY LEGACY 2 IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
MDR Report Key10934530
Report Number3001617766-2020-08438
Device Sequence Number1462497
Product Code DZE
UDI-Device Identifier10841307107576
UDI-Public10841307107576
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K192221
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/21/2024
Device Catalogue Number825713U
Device Lot Number150516
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2020
Initial Report FDA Received Date12/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient Weight103
-
-