| Model Number |
0684-00-0469-01C |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that after approximately eight hours of intra-aortic balloon (iab) therapy, blood was noticed in the tubing.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that after approximately eight hours of intra-aortic balloon (iab) therapy, blood was noticed in the tubing.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Event or Problem Description
|
|
It was reported that after approximately eight hours of intra-aortic balloon (iab) therapy, blood was noticed in the tubing.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and between the catheter and the sheath.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and one leak was detected on the membrane approximately 1cm from the rear seal measuring 0.025cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint # (b)(4).
|
| |
|