| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.(b)(4).A review of the device history record of the product code/lot# combination was conducted with no findings.This report is for the first device reported, for the second device reported that was used on the same patient see mdr 1118880-2020-00310.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during surgery the destination catheter tip was not smooth enough to go through the vessels.There was no patient injury/medical or surgical intervention required.The patient was in stable condition.The procedure outcome was successful as they used the same product to finish the procedure.Additional information was received on 09nov2020.Two devices were used in the case that displayed the sample issue with the edge of the catheter tip not being smooth enough to go through the vessel.There was no reported patient issue as the surgeon was able to successfully use a third device without further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for sheath catheter tip mechanical damage per the information received.Based on the information given, the exact root cause of the event cannot be determined, although it is likely that the sheath tip was damaged due to high resistance.There is no indication that any manufacturing, design or quality system issues may have led to this event.Currently, no action is recommended since this risk evaluation is within the predetermined limits in the fmea.
|
| |
|
Additional Manufacturer Narrative
|
|
One 6fr 45cm destination sheath was returned for product evaluation.The sheath was mated with dilator when received.Visual inspection revealed that damage was observed on the tip of the sheath.Microscopic analysis confirmed that the sheath tip was damaged in a fashion where there was a wrinkle and a tear on the tip of the sheath.Dimensional testing was performed.The outer diameter of the sheath was measured to be 0.1090" which was within specification of 0.1090" ± 0.0030.The inner diameter of the sheath tip was 0.085"which was in the specification of 0.087 " ± 0.002".All measurements were within the manufacturer's specifications.The dilator was subjected to visual analysis and no damage was noted to the dilator.The complaint can be confirmed for sheath tip mechanical damage.Based on the damage observed, it is likely that the damage on the sheath tip can be attributed to difficulties at the point of insertion, possible due to calcification.It is likely that the tip of the sheath came in contact with calcification that propagated various forces that deformed the tip of the sheath.There is no indication that any manufacturing, design or quality system issues may have led to this event.Currently, no action is recommended since this risk evaluation is within the predetermined limits in the fmea.
|
| |
|