BIOMET 3I FULL OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number IFNT411
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); Unspecified Infection (1930); Inflammation (1932)
Date of Event 03/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Patient weight is not provided / unknown.Initial reporter¿s title is not provided / unknown.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reports that the ifnt411 implant lost integration due to peri implantitis and was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10935713
Report Number0001038806-2020-01952
Device Sequence Number19059838
Product Code DZE
UDI-Device Identifier00844868005018
UDI-Public(01)00844868005018(17)220620(10)2017060451(241)IFNT411
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/20/2022
Device Model NumberIFNT411
Device Catalogue NumberIFNT411
Device Lot Number2017060451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2020
Initial Report FDA Received Date12/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
-
-