DEPUY FRANCE SAS - 3003895575 DXTEND CTA HEAD D52 X 26MM; DELTA XTEND IMPLANTS : SHOULDER HUMERAL HEADS Back to Search Results
Model Number 130752026
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient's shoulder was revised due to infection.All implants were removed and an abx spacer was implanted.No surgical delay reported.Doi: (b)(6) 2020, dor: (b)(6) 2020, unknown shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND CTA HEAD D52 X 26MM
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER HUMERAL HEADS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10935737
Report Number1818910-2020-25954
Device Sequence Number4295271
Product Code KWS
UDI-Device Identifier10603295027720
UDI-Public10603295027720
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
MM 49-12
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number130752026
Device Catalogue Number130752026
Device Lot Number5291184
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/13/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date12/02/2020
Supplement Report FDA Received Date01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND CTA HEAD D52 X 26MM; DXTEND HUMERAL SPACER +9MM; DXTEND MOD CENT EPI 1 HA; GLOBAL UNITE REV STEM SZ 12; DXTEND CTA HEAD D52 X 26MM; DXTEND HUMERAL SPACER +9MM; DXTEND MOD CENT EPI 1 HA; GLOBAL UNITE REV STEM SZ 12
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-