WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE FOR 4.5MM VA-LCP CONDYLAR PLATE; CONDYLAR PLATE FIXATION IMPLANT Back to Search Results
Model Number 03.231.001
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on (b)(6) 2020, an aiming arm left not lining up with plate holes when they have a plate attached to the insertion handle, aiming arm right not lining up with plate holes when they have a plate attached to the insertion handle, and insertion handle seems to be out of sink with both aiming arms.It is was discovered during an inspection.There was no patient involvement this complaint involves three (3) devices.This is report 3 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 03.231.001, lot 3748096: manufacturing site: hägendorf.Release to warehouse date: april 26, 2011.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: a product investigation was completed: upon visual inspection, it is observed that the device shows normal wear consistent with the device use which would not contribute to the complaint condition.The etch on the device started to fade and hardly visible.The insertion handle was assembled and disassembled with both right and left aiming arms without any issues.But complete functional testing of the received device was not performed as the mating plates were not returned.Dimensional inspection of the received device was not performed as no physical damage was evident that could contribute to the complaint condition.The relevant drawings were reviewed during the investigation; no design issues or discrepancies were noted during the investigation.The complaint condition cannot be confirmed.A definitive root cause could not be determined for the reported problem.There were no issues during the manufacture of this product that would contribute to this complaint condition.During the investigation, no product design issues, or discrepancies were observed that may have contributed to the complaint condition.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTION HANDLE FOR 4.5MM VA-LCP CONDYLAR PLATE
Common Device Name
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10935804
Report Number2939274-2020-05414
Device Sequence Number871901
Product Code JDP
UDI-Device Identifier10886982077288
UDI-Public(01)10886982077288
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.231.001
Device Catalogue Number03.231.001
Device Lot Number3748096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Initial Date Received by Manufacturer 11/05/2020
Supplement Date Received by Manufacturer12/17/2020
Initial Report FDA Received Date12/02/2020
Supplement Report FDA Received Date12/29/2020
Patient Sequence Number1
-
-