Model Number 82700
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: a used trima access/draw line was returned for investigation.Visual inspection revealed a leak at the ac line bond socket of the 3-1 manifold of the access/draw line.The leak is as a result of a bonding leak path inside the bond socket.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that during a plasmapheresis procedure they found that the set was defective.During plasma collection, they received a centrifuge pressure high alarm and the customer followed prompts on trima to assess possible issues.The customer found that there was air being pulled into the lines just above the manifold where the single tubing from the donor to the manifold connect.There was also blood dripping from this same place.The collection was discontinued without rinseback.A diagram from customer indicated the leak was at the 3-1 manifold.Per the customer no serious injury occured and no medical intervention was required for this event.Patient information and outcome are not available at this time.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in a.1, a.2, a.3, a.4 and b.5.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
Full patient identifier: (b)(6) per the customer no air was returned to the donor.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in b.5, h.6 and h.10 investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.A disposable complaint history search for lot 2009112130 found no other reports of a similar failure.Investigation is in process.A follow up report will be provided.
 
Event or Problem Description
Per the customer, there were no adverse consequences to the donor.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Root cause : the infusion of air into the 3:1 manifold occurred due to the leak.A definitive root cause for the leak could not be determined.Possible causes include but are not limited to: - inadequate contact tubing with a bond socket after solvent application - insufficient application of solvent to the tubing - inadequate insertion of tubing into the bond socket correction: terumo bct has implemented a correction for this incident.Manufacturing staff were made aware of this issue and retrained to the appropriate procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL MULTIPLASMA, ISBT
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key10935874
Report Number1722028-2020-00530
Device Sequence Number11555934
Product Code GKT
UDI-Device Identifier05020583827004
UDI-Public05020583827004
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2022
Device Model Number82700
Device Catalogue Number82700
Device Lot Number2009112130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2020
Supplement Date Received by Manufacturer12/09/2020
02/17/2021
03/15/2021
Initial Report FDA Received Date12/02/2020
Supplement Report FDA Received Date12/22/2020
03/01/2021
03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age00064 YR
Patient Weight132
-
-