• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC HIGH PRESSURE BRAIDED CONNECTING TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC HIGH PRESSURE BRAIDED CONNECTING TUBE Back to Search Results
Model Number G00060
Date of Event 11/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Common name & product code: dtl adaptor, stopcock, manifold, fitting, cardiopulmonary bypass.(b)(6).Pma/510(k) number: exempt.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, prior to use during an unknown procedure, a high pressure braided connecting tube leaked at the connection site.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No new patient or event information to report.
 
Additional Manufacturer Narrative
Description of event: as reported, prior to use during an unknown procedure, a high pressure braided connecting tube leaked at the connection site.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation ¿ evaluation.A document based investigation was performed including a review of complaint history, device history record, documentation, manufacturing instructions, quality control data, and specifications.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was made out of specification.A search of cook¿s inventory determined that all devices from the complaint lot were already out of cook¿s control and thus a representative device from the lot could not be examined.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.There is no ifu provided with this device.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.It is most likely that the device was inadvertently damaged during either shipping or storage.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH PRESSURE BRAIDED CONNECTING TUBE
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10936318
Report Number1820334-2020-02221
Device Sequence Number11449293
Product Code DTL
UDI-Device Identifier00827002000605
UDI-Public(01)00827002000605(17)230304(10)13027856
Combination Product (Y/N)N
Initial Reporter CountryKR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2023
Device Model NumberG00060
Device Catalogue NumberHPCT8.8-120-M-FM
Device Lot Number13027856
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/25/2020
Supplement Date Received by Manufacturer03/11/2021
Initial Report FDA Received Date12/02/2020
Supplement Report FDA Received Date03/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-