STRYKER ENDOSCOPY-SAN JOSE PACKAGING, FORMULA SHAVER (HAND CONTROL); BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number 0375704500
Medical Device Problem Code Insufficient Cooling (1130)
Health Effect - Clinical Codes Burn(s) (1757); Burn, Thermal (2530)
Date of Event 11/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the patient was burned.
 
Additional Manufacturer Narrative
Alleged failure: shaver over heating burned salesforce case number (b)(4).Harm: burned pt.Delay: 5m.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) excessive force applied to the cutter/burr by the user with poor or no suction during use.(excessive force may cause friction due to bur shaft assembly and housing assembly interaction.) (2) tissue blocking the suction pathway would prevent hot fluid from being removed from the joint.(3) poor arthroscopy pump suction.(4) additionally, during arthroscopic procedures, it is common to use an rf energy wand.Inherent to use of an rf probe is the risk of potential superficial patient burns due to the heat generated from the rf output.Therefore use of the rf probe can also be considered a possible root cause as well.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
 
Event or Problem Description
It was reported that the patient was burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGING, FORMULA SHAVER (HAND CONTROL)
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10936986
Report Number0002936485-2020-00503
Device Sequence Number702565
Product Code GFA
UDI-Device Identifier07613327057638
UDI-Public07613327057638
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0375704500
Device Catalogue Number0375704500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2020
Initial Date Received by Manufacturer 11/05/2020
Supplement Date Received by Manufacturer11/05/2020
Initial Report FDA Received Date12/02/2020
Supplement Report FDA Received Date01/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-