| Model Number |
0375704500 |
| Medical Device Problem Code |
Insufficient Cooling (1130)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Burn, Thermal (2530)
|
| Date of Event |
11/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was burned.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: shaver over heating burned salesforce case number (b)(4).Harm: burned pt.Delay: 5m.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) excessive force applied to the cutter/burr by the user with poor or no suction during use.(excessive force may cause friction due to bur shaft assembly and housing assembly interaction.) (2) tissue blocking the suction pathway would prevent hot fluid from being removed from the joint.(3) poor arthroscopy pump suction.(4) additionally, during arthroscopic procedures, it is common to use an rf energy wand.Inherent to use of an rf probe is the risk of potential superficial patient burns due to the heat generated from the rf output.Therefore use of the rf probe can also be considered a possible root cause as well.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was burned.
|
| |
|