MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ5 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ5 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-20-507
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Muscular Rigidity (1968); Pain (1994)
Date of Event 03/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient underwent a right knee revision due to pain, stiffness, nickel allergy, adhesions, and tibial tray loosening at the cement to implant interface.Competitor cement was used during the primary operation.The patellar component was retained.Competitor products were implanted with competitor cement.There is no indicated intra-operative complications.Doi: (b)(6) 2017; dor: (b)(6) 2019; right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ5 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10937348
Report Number1818910-2020-26011
Device Sequence Number790909
Product Code JWH
UDI-Device Identifier10603295045663
UDI-Public10603295045663
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-20-507
Device Catalogue Number151620507
Device Lot NumberH18133
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/17/2020
Supplement Date Received by Manufacturer01/15/2021
Initial Report FDA Received Date12/03/2020
Supplement Report FDA Received Date01/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ5 7MM; ATTUNE CR FEM RT SZ 5 NAR CEM; ATTUNE FB TIB BASE SZ 4 CEM; ATTUNE MEDIAL DOME PAT 35MM; COMPETITOR ZIMMER PALACOS BONE CEMENT; COMPETITOR ZIMMER PALACOS BONE CEMENT; ATTUNE CR FB INSRT SZ5 7MM; ATTUNE CR FEM RT SZ 5 NAR CEM; ATTUNE FB TIB BASE SZ 4 CEM; ATTUNE MEDIAL DOME PAT 35MM; COMPETITOR ZIMMER PALACOS BONE CEMENT; COMPETITOR ZIMMER PALACOS BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-