BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 11/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.While a 2.75mm x 15mm quantum maverick balloon catheter was being unpacked, the shaft broke.The procedure was completed with another of the same device.
 
Event or Problem Description
It was reported that shaft break occurred.While a 2.75mm x 15mm quantum maverick balloon catheter was being unpacked, the shaft broke.The procedure was completed with another of the same device.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a quantum maverick mr balloon catheter.The balloon was tightly folded.Analysis of the tip, balloon, inner/outer shaft and hypotube included microscopic and visual inspection.Inspection revealed the hypotube fractured 98.1cm from the tip of the device, and there were numerous kinks in the hypotube.The fracture faces were ovalized, suggesting the area had been kinked prior to separating.Inspection of the rest of the device found no other damage or defects.The reported shaft fracture was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10938107
Report Number2134265-2020-16848
Device Sequence Number3190129
Product Code LOX
UDI-Device Identifier08714729392774
UDI-Public08714729392774
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2023
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0025332956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Initial Date Received by Manufacturer 11/17/2020
Supplement Date Received by Manufacturer01/14/2021
Initial Report FDA Received Date12/03/2020
Supplement Report FDA Received Date02/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight55
-
-