| Model Number |
SMPX755 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
11/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient who had a 755cc mentor memorygel breast implant placed experienced right sided baker grade iv capsular contracture post procedure.As a result, replacement with mentor gel was performed on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
The impacted product was received by the failure analysis lab on (b)(6) 2021.The evaluation of the returned device was completed by the failure analysis lab on (b)(6) 2021.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|