SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE M 12.9X12.9 CTN10; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 66801068
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the silicon adhesion is coming off onto the plastic backing.The product has not been submitted to a customer or used on a patient.
 
Additional Manufacturer Narrative
H3, h6: we have now concluded our investigation into this complaint.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.The complaint history file contains further instances/ related events of the reported event.The device, intended for use in treatment, was not returned for evaluation.However a relationship between the reported event and the device was confirmed from images provided.The wound contact surface of allevyn life is coated with a gentle silicone adhesive layer that ensures non-traumatic removal at dressing changes.The silicone adhesive will feel sticky when compared to an acrylic adhesive and is specially designed to be very soft and gentle, but can soften further, particularly at higher temperatures.This is an inherent characteristic of all of the silicone adhesives and gives them their soft / gentle properties.It is therefore possible if the product has been stored at a high temperature, this could have contributed to the reported issue.An ongoing internal project is being carried out into this failure mode, therefore no further actions are deemed necessary into this specific complaint.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.We will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN LIFE M 12.9X12.9 CTN10
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10939160
Report Number8043484-2020-04042
Device Sequence Number4329697
Product Code NAD
UDI-Device Identifier05000223478018
UDI-Public05000223478018
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66801068
Device Catalogue Number66801068
Device Lot Number202009
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date12/03/2020
Supplement Report FDA Received Date01/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-